MATRIX and MATRIX-PLUS Portable Gas Detector
Medical Gas Verification – Made Simple
A Multi-Sensor Transportable Analyzer for Medical Gas Verifier/Certifier Field Usage
The MATRIX medical gas verifier is designed for the real-time analysis and verification of hospital gases to ensure compliance with NFPA 99.
The MATRIX offers the operator two key advantages:
- Reduces the time to complete a full analysis on station.
- Avoids the complexity of using single purpose instruments.
The MATRIX uses an (8) channel multi-sensor system to simultaneously monitor the following gases:
- Carbon Monoxide
- Carbon Dioxide
- Oxygen
- Methane
- Nitrous Oxide
- Trace Hydrocarbons
- Anesthesia Gases
- Dew Point
The MATRIX is a transportable instrument that is fully self-contained and able to operate on-site for 12 hours with no external support requirements. The real-time analysis data is displayed on the front LCD panels, which have integral alarm concentration indicators and loss of flow alarms.
The MATRIX Plus offers a data logging option which allows the analysis data to be displayed on a PC via a USB connection. This permits the real-time data to be displayed in a graphical plot. The data log can be archived with a location file name for use in reports. Additionally, the MATRIX Plus can be configured with a sampling pump for those applications which are not pressurized.
- Real-time monitoring for 8 gases
- Faster station analysis
- Easily transportable and durable case
- Battery operated (12 hrs)
- Optional datalogger
- Complete field analysis in one instrument
Display: Alpha numeric LCD with Backlight
Alarm: LED visual indicator per channel and flow alarm
Operating Temperature: 32º to 77ºF (0º to 25ºC) recommended
Humidity Range: 0-95% (non-condensing)
Power: Rechargeable lithium battery & AC power cord
Run Time: Up to 12 hours on a fully charged battery, or continuous with AC power cord
Inlet Pressure for Dew Point: 55 psig (user provided regulator)
Inlet Flow Rate for Sensors:0.5 liters/minute
Enclosure: Polypropylene-copolymer case
Size: 19 x 15 x 8 inches (48 x 38 x 20 cm)
Weight: 19 lb. (8.6 kg)
SPECIFICATIONS FOR GASES MONITORED*
GAS | RANGE | ALARMPOINT1 | ALARMPOINT2 | ACCURACY AT ALARM POINTS | SENSOR TYPE | DISPLAY INCREMENT |
1.Carbon Monoxide(CO) | 0-50ppm | 10ppm | 20ppm | ±1ppm | EC | 1ppm |
2.Carbon Dioxide (CO2) | 0-5000ppm | 1000ppm | 2000ppm | ±20ppm | NDIR | 10ppm |
3.DewPoint | -112°to+68°F | +35°Fat55PSIG | +50°F at 55 PSIG | ±2°F | TFP | 1°F or 1°C |
4. Trace Hydrocarbons | 0-20ppm | 5ppm | 10ppm | ±1ppm | PID | 0.1ppm |
5.Anesthesia Gases | 0-100ppm | 50ppm | 100ppm | ±5ppm | MOS | 1ppm |
6. Oxygen(O2) | 0-100%byVol. | ↓95%byVol. | ↓90%byVol. | ±0.1%byVol. | PP/EC | 0.1%byVol. |
7. Methane(CH4) | 0-5000ppm | 500ppm | 1000ppm | ±10ppm | NDIR | 10ppm |
8.Nitrous Oxide(N2O) | 0-100%byVol. | ↓95%byVol. | ↓90%byVol. | ±3%byVol. | NDIR | 1%byVol. |
9.Carbon Dioxide(CO2) | 0-100%byVol. | ↓95%byVol. | ↓90%byVol. | ±3%byVol. | NDIR | 1%byVol. |
*NOTE: Maximum 8 gases per instrument
- Anesthesia Gaes
- Carbon Dioxide (CO2)
- Carbon monoxide (CO)
- Dew Point (DP)
- Hydrocarbons
- Methane (CH4)
- Nitrous oxide (N2O)
- Oxygen (O2)
- Water/Dew Point
Options:
The MATRIX-PLUS offers a datalogging option which allows the analysis data to be displayed on a PC via a USB connection, and can be configured with a sampling pump for those applications which are not pressurized.
To place an order, please contact our experienced and professional customer service team:
PHONE: 734-761-1270
EMAIL: [email protected]
If you cannot find what you need or require a custom part, simply call and speak to one of our knowledgeable gas monitoring experts. They will be able to walk you through the ordering process quickly and professionally.